Cobas IT Middleware Core Unit (2015-04-01) — recall 53416
Health Canada — Medical Devices · Canada · 2015-04-29 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Cobas IT Middleware feature "sample closing" could fail if a "rebuilt sample" message was sent by the host before the "sample closing date" already established by the user. The sample result mismatch may occur, if the customer allows recycling patient's ids and if their system was set to allow "sample rebuild".
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.