Hamilton Medical Infant Flow Sensor (2015-05-12) — recall 53247

Health Canada — Medical Devices · Canada · 2015-06-03 · Type I

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
53247
Date of alert
2015-06-03
Country / jurisdiction
Canada
Risk level
Type I
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Under some operating conditions, it may be possible that water from the humidifier may condensate within the flow sensor measurement chamber and affect the flow measurement, resulting in more volume being measured than being delivered to the patient. The Instructions for Use is being revised to inform the user of the correct position to have the flow sensor rest at.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/hamilton-medical-infant-flow-s

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.