Product Wiki · Recalls · Health Canada — Medical Devices

Surgical Light SA Ceiling Suspension (2015-04-29) — recall 53236

Health Canada — Medical Devices · Canada · 2015-06-01 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
53236
Date of alert
2015-06-01
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: There is a potential defect in the ceiling SA suspension arms of the Maquet Surgical Lights model powerLED, Hled, Xten, Standop Volista and Hanaulux. Maquet received some complaints of grinding on double or triple arm suspensions with paint possibly chipping from the junction between two arms. Paint chips or particles can fall down from the suspension. This may happen during surgical procedures and paint chips may fall inside the surgical field of a patient.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/surgical-light-sa-ceiling-susp

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.