Product Wiki · Recalls · Health Canada — Medical Devices

Intellivue Information centre IX - software (2015-04-13) — recall 53116

Health Canada — Medical Devices · Canada · 2015-05-08 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
53116
Date of alert
2015-05-08
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Philips has become aware that when HL7 lab messages are sent to a PIIC iX system that has been localized for comma delimiters, the lab result is incorrectly displayed on the bedside monitor. An expected delimiter is stripped from the lab value and the number shown is of a higher magnitude than expected. If lab results are incorrect, a clinician may initiate incorrect treatment or delay necessary treatment this problem only happens when the PIIC iX is configured during set up for regional settings where the delimiter is a comma. The lab option must be in use, and the workflow utilizes the automatic lab interface displayed at the bedside monitor.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/intellivue-information-centre-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.