Product Wiki · Recalls · Health Canada — Medical Devices

Focus Dental X-ray System (2014-01-14) — recall 53000

Health Canada — Medical Devices · Canada · 2014-12-18 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
53000
Date of alert
2014-12-18
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: An internal investigation at Palodex revealed that wall mounting units that were missing structural welds had been delivered to Palodex by the supplier. It was determined that the defective wall mounting units could potentially cause a safety hazard, if the top weld is missing and wall mounting unit fails to support the device.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/focus-dental-x-ray-system-2014

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.