ADVIA Centaur and Centaur CP Systems (2015-01-27) — recall 52791
Health Canada — Medical Devices · Canada · 2015-02-06 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: There is a high bias on results greater than 30 ng/ml (95.4 nmol/l) when compared to the master curve when using select ADVIA centaur progesterone kit lots. The high bias may cause neat serum samples with results greater than 60 ng/ml (190.8 nmol/l) to recover within the assay range upon dilution (dilution factor applied). Samples with values below 30 ng/ml (95.4 nmol/l) do not demonstrate this bias.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.