Reported risk: This recall has been initiated due to the potential that, it was observed that creatinine rate blank method calibrator and control results were reporting values lower than target values indicated in the associated ABX Pentra Multical, N control and P control annexes. Horiba has reassigned the target values and updated annexes for the products and lot numbers.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.