Reported risk: The affected CO 2 analyzer modules may experience a problem at the end of the calibration of the Capnostat 3 Sensor, causing the calibration to fail and activate the medium priority alarm "CO 2 module error" on the user interface of the Servo Ventilator. This issue only affects the optional CO 2 analyzer functionality, preventing it from displaying any measured CO 2 values. It has no effect on any other functionality of the ventilator.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/servo-ventilator-system-co2-an

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.