Product Wiki · Recalls · Health Canada — Medical Devices

VWR Pre-Calibrated Microhematocrit Tubes With Heparin (2014-08-11) — recall 52645

Health Canada — Medical Devices · Canada · 2015-01-23 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
52645
Date of alert
2015-01-23
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Drummond manufactures two products under the name of pre-cal. They are similar in appearance but have different OD's and ID's. In this instance the machine operator misread the specification and used the larger ID glass. Both of these tubes will deliver the same results when performing a hematocrit test. The pre-calibrated tube is used to determine a more consistent column height of blood at 60mm.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/vwr-pre-calibrated-microhemato

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.