Health Canada — Medical Devices · Canada · 2015-01-14 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: During internal stability testing of the reconstituted PreciControl Multimaker of lot 178364, a reduced stability of PIGF became apparent. Whereas the stability claim for all analytes in PreciControl Multimaker is 72 hours at 2-8c, this claim does not apply for PIGF As a consequence, the stability of PIGF has been reduced accordingly in the method sheet of the PreciControl Multimaker. The claim at 2-8c was removed but the frozen and ambient temperature stability remained the same as they are not affected.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.