ACROBAT V Vacuum Off-Pump System (2014-08-18) — recall 52411
Health Canada — Medical Devices · Canada · 2014-09-16 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: An internal finding identified the presence of pinholes/cuts in some of the Tyvek pouches of the vacuum tubing sets included as an accessory to these products. As a result, there is a potential for the pouch integrity to be compromised, which may result in compromised sterility of the product. If the compromised product is used in surgery there is a potential risk of infection for the patient.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.