Dimension Vista System - MMB Calibrator (2014-08-18) — recall 52333
Health Canada — Medical Devices · Canada · 2014-08-25 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Siemens healthcare diagnostics has determined that Dimension Vista Mass CKMB Isoenzyme Calibrator (MMB Cal) shows drift during routine stability monitoring that exceeds the acceptance criteria for drift over the shelf life of this product (12 months). Siemens has observed drift from 3% to 12% at MMB concentrations across the assay range. Depending on quality control limits, this drift may not have been detected. Due to the drift, Siemens is revising the expiration date of this product (to 6 months). Siemens is issuing an urgent field safety notice letter for the aforementioned lots of Dimension Vista MMB Cal.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.