Joerns Healthcare EasyCare Bed (2014-10-07) — recall 52194

Health Canada — Medical Devices · Canada · 2014-10-29 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
52194
Date of alert
2014-10-29
Country / jurisdiction
Canada
Authority classification
Type II
Notice type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Joerns healthcare has identified an issue with EasyCare Beds configured w/the care-lock system where in some instances the care-lock system, after being engaged by stepping on the red or right side pedal, through normal daily use and patient/care-giver interaction with the bed can disengage i.e., revert to a non-locked state. The care-lock system is located at the foot end of the bed, there are also steering locks located on the dual wheel casters at the head of the bed. This does not affect the steering locks located on the dual wheel casters at the head of the bed.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/joerns-healthcare-easycare-bed

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.