IntelliVue MP5 and MP5SC Patient Monitor System and IntelliVue… — recall 52165
Health Canada — Medical Devices · Canada · 2014-10-14 · Type II
As described by Health Canada — Medical Devices: IntelliVue MP5 and MP5SC Patient Monitor System and IntelliVue Patient Monitor MP5T (2014-09-24)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Philips has become aware that in time-synchronized automatic/sequence mode the NBP automatic measurement series is stopped inadvertently, if the repetition interval is increased manually by the user (e.g. From 5 to 15 minutes) in between two consecutive measurements: a prompt message "any ongoing NBP measurement and automatic cycle stopped" is displayed and a single tone sounds when the measurement series is stopped. After the automated measurement series has been stopped, NBP numeric, repetition time, and time to next measurement are no longer displayed, i.e. The NBP numeric field is blank.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.