Steris 5085 and 5085SRT Surgical Table (2014-10-13) — recall 52148
Health Canada — Medical Devices · Canada · 2014-10-22 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Steris has learned through customer feedback and field service experience that the shroud cover for the hydraulic column may become misaligned and damaged as a result of extreme table articulation or excessive pressure on the side or base of the table. In most instances, the shroud damage is minor and does not affect the vertical up/down movement of the hydraulic column. In cases of severe shroud damage, the shroud sections may become misaligned such that vertical movement of the hydraulic column is impeded.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.