Health Canada — Medical Devices · Canada · 2014-10-09 · Type II
As described by Health Canada — Medical Devices: UniCel DxC 600I Synchron Access Clinical Analyzer and Access 2 Immunoassay Systems (2014-09-28)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Beckman coulter has determined that the Access 2 console PC (B23083) may experience an "MFC exception" error during normal operation. If this error occurs, the system will exit the Access 2 user interface (UI) screen and display a window that notifies the user of a "0==I tubes can count" event. If the UI is re-launched, the system will prevent access to the sample manager screen to request samples for processing and the same "MFC exception" error will occur.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.