Valve Tubing 841-797 for ABL90 Flex Analyzer (2014-09-11) — recall 52090

Health Canada — Medical Devices · Canada · 2014-10-09 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
52090
Date of alert
2014-10-09
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Radiometer has recently become aware that the valve tubing, 841-797, for the ABL90 Flex Analyzer may potentially leak at the steel tube connector in the hemolyzer unit. This may result in droplets of fluid dripping onto the plate underneath the valve or onto the bottom of the OXI module housing.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/valve-tubing-841-797-abl90-fle

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.