Product Wiki · Recalls · Health Canada — Medical Devices

Columbia Agar with 5% Sheep Blood and Trypticase Soy Agar with 5%… — recall 52053

Health Canada — Medical Devices · Canada · 2014-08-22 · Type II

As described by Health Canada — Medical Devices: Columbia Agar with 5% Sheep Blood and Trypticase Soy Agar with 5% Sheep Blood (TSA II) / MacConkey II Agar (2014-07-10)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
52053
Date of alert
2014-08-22
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: BD has recently confirmed a low level of surface and subsurface contamination of Listeria monocytogenes on the prepared plated media. May be hazardous to health. The contamination could be erroneously reported as an actual patient result causing misdiagnosis and/or mistreatment of a non-serious illness.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/columbia-agar-5-sheep-blood-an

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.