Health Canada — Medical Devices · Canada · 2014-07-03 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: During testing shaver hand piece tray failed Sterrad 100s half cycle testing and did not demonstrate a sterility assurance level of 10-6 for the TPS shaver. It has been determined that the reprocessing guide (P10299 Rev. A) is the non-conforming part of the shaver hand piece tray and the sterilization method with inadequate sterilization validation (Sterrad 100s) needs to be removed as a sterilization option for the TPS small joint shaver.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.