Product Wiki · Recalls · Health Canada — Medical Devices

Intouch Bed Caster Stem (2014-07-02) — recall 51674

Health Canada — Medical Devices · Canada · 2014-07-22 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
51674
Date of alert
2014-07-22
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Stryker received complaints which allege that a caster has broken on the top of the caster stem where the brake rod inserts into the caster. If more than one caster on a unit is damaged, brake functions may be affected and result in reduced brake holding force. If only one caster on the bed is damaged, brake functionality continues to meet specifications.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/intouch-bed-caster-stem-2014-0

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.