Progressa Bed Mattress (2014-03-19) — recall 51528
Health Canada — Medical Devices · Canada · 2014-04-14 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Hill-Rom has become aware that certain Progressa beds with integrated mattresses may have a potential issue during turn assist mode. An error code 31C9 may be displayed on the graphical user interface and the wrench light will turn on indicating service required. The 31C9 error means the turn assist bladder is unable to hold the required pressure to operate as intended.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.