Discovery MR750 and MR450 MR Systems and Optima MR450W MRI Systems… — recall 51527

Health Canada — Medical Devices · Canada · 2014-04-14 · Type II

As described by Health Canada — Medical Devices: Discovery MR750 and MR450 MR Systems and Optima MR450W MRI Systems (2014-03-28)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
51527
Date of alert
2014-04-14
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: The gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance may lead to increasing heat during scanning, which, in turn, can cause material in the rear endbell enclosure to overheat, generate smoke in the scan room and cause some localized melting in the endbell.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/discovery-mr750-and-mr450-mr-s

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.