Steri-Vac Gas Sterillizer/Aerator (2014-03-17) — recall 51490
Health Canada — Medical Devices · Canada · 2014-04-14 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: 3M received a single report of a potential exposure OT ethylene oxide (EO) at a hospital installation outside of Canada. The cause was traced to a rare situation involving a failure of the EO cartridge puncture assembly part manufactured between November 10, 2006 and December 14, 2007.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.