Product Wiki · Recalls · Health Canada — Medical Devices

Aerial Cushion (2014-02-13) — recall 51469

Health Canada — Medical Devices · Canada · 2014-04-14 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
51469
Date of alert
2014-04-14
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Recent routine internal fire testing to test flammability after multiple washings of the device (N=5) of the whole cushion (cover and foam) in water after 40 degrees celsius, revealed a potential risk of flammability from the product (the "testing failure"). This testing failure contradicts the information contained in the company's instructions for use wherein it states: "hand wash all foam breathair components in warm water at 40 degrees celsius using a standard mild detergent. Do not use fabric softner. Rinse well."
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/aerial-cushion-2014-02-13

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.