Heartstart MRX Defibrillator / Monitor and QCPR-CPR Meter Upgrade Kit… — recall 51391
Health Canada — Medical Devices · Canada · 2014-03-25 · Type II
As described by Health Canada — Medical Devices: Heartstart MRX Defibrillator / Monitor and QCPR-CPR Meter Upgrade Kit (2014-02-26)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Philips has become aware of the following issues regarding the Heartstart MRX monitor/defibrillator: when the Heartstart MRX is in AED mode, the Q-CPR meter is designed to display the 'Do not touch the patient' icon during the shock advisory analysis period. If the user continues to deliver chest compressions when the shock advisory analysis begins, the 'Do not touch the patient' icon may incorrectly remain on the screen after analysis or shock delivery is complete. When the Heartstart MRX is in manual mode, the Q-CPR meter is designed to display the 'Do not touch the patient' icon when the device is charging to the selected energy. If the clinician continues to provide chest compressions while the device is charging, the Q-CPR meter 'Do not touch the patient" icon may incorrectly remain on the screen after shock delivery is complete.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.