Heartstart MRX Defibrillator / Monitor (2014-02-24) — recall 51368

Health Canada — Medical Devices · Canada · 2014-03-25 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
51368
Date of alert
2014-03-25
Country / jurisdiction
Canada
Authority classification
Type II
Notice type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Philips has become aware that when used in hospital transport and pre-hospital (EMS) environments, the connection between the pads/paddles therapy cable (including pads CPR therapy cable, external paddles cable, and internal paddles cable/adapter) and the MRX therapy connection port could experience high levels of stress causing accelerated wear. Wear of the connection could prevent the device from sensing that the pads/paddles therapy cable is connected. This wear also could cause the MRX to inappropriately identify the pads therapy cable, external paddles, or internal paddles. Pads/paddles therapy cable and therapy connection port wear could pose a risk for patients and/or caregivers.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/heartstart-mrx-defibrillator-m

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.