Health Canada — Medical Devices · Canada · 2014-04-04 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: When using 54340D Allset+ Gold HLA ABC Low Resolution kit lot number 038 1249868, Primer mix PM085C has been shown to produce a weak false positive reaction of correct size (420 base pair PCR Amplicon) in the presence of samples containing C*04 homozygote allele(s). Primer mix PM085C can be found in lane 78. As a result of the false positive reaction, samples containing two C*04 alleles could produce a mistype or no-type result depending on the reaction pattern corresponding to the full ABC HLA typing.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.