Advia Centaur, Centaur CP, and Centaur XP Systems — recall 51219
Health Canada — Medical Devices · Canada · 2014-04-04 · Type III
As described by Health Canada — Medical Devices: Advia Centaur, Centaur CP, and Centaur XP Systems - Procaclitonin Assay (PCT) (2014-02-21)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Siemens Healthcare Diagnostics received customer feedback regarding the clinical utility of Advia Centaur Brahms Procalcitonin (PCT) quality control level 1. Quality control level 1 currently has a range of plus or minus 50% of the target. Siemens internal investigation confirmed that this range was incorrectly set. Siemens has revised the range of Advia Centaur Brahms Procalcitonin (PCT) quality control level 1 to plus or minus 30% of the target. A review of test data confirmed the assay is capable of achieving a narrower range of plus or minus 30% of the target.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.