Engen Automated Sample Handling System (2014-04-11) — recall 51148
Health Canada — Medical Devices · Canada · 2014-04-30 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: In Engen systems with TC-Automation (TCA) software version 2.6, version 3.2 and version 3.5 configured with bypass modules for Advia Centaur, Abbott Architect, Vitros and/or Tosoh Aia systems, a sample tube carrier being released in the bypass module may not be detected and, therefore, the next sample to be metered may not advance for processing. No error message will be posted if this occurs. The bypass module will stop processing samples indefinitley until an operator detects the issue and intervenes.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.