Product Wiki · Recalls · Health Canada — Medical Devices

High Frequency Oscillatory Ventilators Models 3100A and 3100B… — recall 50885

Health Canada — Medical Devices · Canada · 2013-12-17 · Type II

As described by Health Canada — Medical Devices: High Frequency Oscillatory Ventilators Models 3100A and 3100B (2013-03-11)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
50885
Date of alert
2013-12-17
Country / jurisdiction
Canada
Authority classification
Type II
Notice type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: There have been reports of failures of Cap / Diaphragm assemblies. Specific lots of Cap / Diaphragm assemblies used with 3100A or 3100B high frequency oscillatory ventillators could fail thereby preventing the targeted mean airway pressure (39 - 43 cm H20) being reached during system calibration test set up prior to being used on patients.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/high-frequency-oscillatory-ven

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.