Product Wiki · Recalls · Health Canada — Medical Devices

Advia Chemistry Systems (2014-01-23) — recall 50825

Health Canada — Medical Devices · Canada · 2014-03-19 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
50825
Date of alert
2014-03-19
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Siemens Healthcare Diagnostics is conducting a recall for Advia 1800 and Advia 2400 ISE firmware v2.23. Advia Chemistry Chloride Electrodes may fail prematurely during calibration due to a chloride bias parameter error. The operator will receive an ng flag which requires immediate action to be taken on the calibration samples.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/advia-chemistry-systems-2014-0

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.