Abbott FreeStyle glucose test strips may produce false test results… — recall 50761
Government of Canada recall portal · Canada · 2014-03-11
As described by Government of Canada recall portal: Abbott FreeStyle glucose test strips may produce false test results with certain devices
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Health Canada is informing Canadians that when Abbott FreeStyle glucose test strips are used with the devices listed below, there is a potential for users to receive a lower-than-actual blood sugar reading. False low blood sugar readings may lead to improper treatment decisions for users, such as increasing the consumption of sugar or not taking enough insulin, which in turn could lead to hyperglycemia (elevated blood sugar levels). The three devices are: Abbott Diabetes Care's FreeStyle Blood Glucose Meter and FreeStyle Mini Blood Glucose Meter, and Insulet's OmniPod Insulin Management System. Consumers with the affected FreeStyle meters should contact Abbott Diabetes Care at 1-877-576-9368. Consumers with the Insulet OmniPod Insulin Management System should contact Abbott Diabetes Care at 1-877-607-6364 regarding the Freestyle strips, or if they have questions about the use of the system they should contact Insulet at 1-855-763-4636. The two companies are notifying customers to inform them of this issue. Consumers who have concerns related to the use of these devices should speak with their health care professional.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.