V Series Monitoring System (2014-01-27) — recall 50751
Health Canada — Medical Devices · Canada · 2014-03-18 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: The V dock power pins connecting the V series monitor to the V dock may lose spring tension, potentially causing the screen to temporarily go blank, lose communication with the central station (when one is in use) or shut down.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.