Unicel DXI 600 and DXI 800 Access Immunoassay Systems (2014-01-29) — recall 50733
Health Canada — Medical Devices · Canada · 2014-03-18 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: The Access Thyroglobulin Calibrators are a lyophilized product and as such need to be reconstituted prior to use. The reconstitution time of 30 minutes as stated in the thyroglobulin calibrators instructions for use (IFU) is insufficient for the four affected lots of thyroglobulin calibrators. This has resulted in high patient and control results that are on average greater than 1 standard deviation high.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.