Dimension Vista System - Enzyme Calibrator (2014-01-20) — recall 50683

Health Canada — Medical Devices · Canada · 2014-02-11 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
50683
Date of alert
2014-02-11
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: There have been reports of quality control (QC) shifts with pseudocholinesterase (PCHE) and imprecision during PCHE calibration with Dimension Vista enzyme 1 calibrator (ENZ 1 cal) lots 3FD034 and 3HD012. Investigation by Siemens confirms a loss of PCHE activity in the affected lots of enzyme 1 calibrator, causing an elevation of qc by as much as 80% and in some instances, imprecision of replicates during calibration.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/dimension-vista-system-enzyme-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.