Immulite 2000 System - Total IgE Assay (2013-12-27) — recall 50633

Health Canada — Medical Devices · Canada · 2014-02-11 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
50633
Date of alert
2014-02-11
Country / jurisdiction
Canada
Authority classification
Type III
Notice type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Siemens Healthcare Diagnostics has confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the Immulite Systems Total LGE assays. The Immulite Total LGE controls (IECA1,2) will not detect the bias.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/immulite-2000-system-total-ige

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.