Progeny Vantage Panoramic X-Ray System with Cephalometric (2013-09-27) — recall 50586

Health Canada — Medical Devices · Canada · 2013-10-28 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
50586
Date of alert
2013-10-28
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: This field action is being initiated to correct a recently identified real time controller (RTC)firmware error found in firmware versions 3.0, 3.1, 3.2, and 3.3. This error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. "failure to stop"). If this occurs while a patient is positioned under the column, there is the potential for injury, such as bruising, sprains, or resulting from quick movements to avoid the moving column.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/progeny-vantage-panoramic-x-ra

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.