Health Canada — Medical Devices · Canada · 2013-10-28 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Two lots of F25ONRe Dialyzer were recalled in February, 2011 (Incident ID 61779) due to increased complaints associated with internal blood leaks. Quality trending revealed that lot 10HU02012 and 10LU02016 demonstrated a higher blood leak rate than for other lots. Following a FDA inspection, a decision was made by the legal manufacturer to place additional 20 lots as they were manufactured under the same production process and could not be definitively segregated from the recalled lots.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.