Product Wiki · Recalls · Health Canada — Medical Devices

Hill Rom 1000, Advanta 2, and Careassist P1170 Hospital Beds… — recall 50449

Health Canada — Medical Devices · Canada · 2013-10-15 · Type II

As described by Health Canada — Medical Devices: Hill Rom 1000, Advanta 2, and Careassist P1170 Hospital Beds (2013-03-28)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
50449
Date of alert
2013-10-15
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: The manufacturer of the power control printed circuit board (PCB) changed the control relays to another manufactures relay and that due to some switch timing issues the electrical current traveling through the relays became large enough that the relay contacts were welded together. Once the relay contacts for a function became welded, that function ceases to operate. The welding of the contacts seems to happen more when more than one function is operated at the same time.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/hill-rom-1000-advanta-2-and-ca

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.