Advia Chemistry Systems (2013-11-11) — recall 50375

Health Canada — Medical Devices · Canada · 2013-12-04 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
50375
Date of alert
2013-12-04
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Siemens confirms that three issues were identified where the system may pass calibration and QC, when it should not. These include auto-calibration with flagged system related errors, auto-calibration with a lot change and reagent pack switching mid-calibration or mid-QC. Users are being advised when any of the three situations occur, to perform the calibration or QC again.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/advia-chemistry-systems-2013-1

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.