Osseocare Pro (2013-10-24) — recall 50374

Health Canada — Medical Devices · Canada · 2013-12-04 · Type I

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
50374
Date of alert
2013-12-04
Country / jurisdiction
Canada
Risk level
Type I
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: When using the Osseocare Pro Ipad application release 1.2.0.7, the grey button located on the foot pedal has no effect on the rotation direction. Pressing the grey button normally sets the motor in reverse rotation mode. However, when pressing the grey button with the software release 1.2.0.7, this does not change the rotation direction. This malfunction may result in the screw tap and/or implant being placed deeper than intended, which in the worst case could damage vital anatomical structures (Mandibular nerve and blood vesseles, or penetration in to the Lumen of maxillary sinus or nasal cavity).
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/osseocare-pro-2013-10-24

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.