Endeavour Light System (2013-10-28) — recall 50248
Health Canada — Medical Devices · Canada · 2013-11-29 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: If user fails to secure the cable to the temple arm of his/her loupes with the spiral temple clip provided with the device, strain on the cable (cable pull forces in excess of 7 pounds) where it interfaces to the headlight may result in an electrical open that prevents power from reaching the headlight LED.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.