Product Wiki · Recalls · Health Canada — Medical Devices

Sapphire Multi-Therapy Infusion Pump (2013-10-08) — recall 50228

Health Canada — Medical Devices · Canada · 2013-11-13 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
50228
Date of alert
2013-11-13
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Under certain cumulative conditions the time spent in each of the steps may deviate from the programmed time for the step. The infusion time will therefore be longer or shorter than the programmed time and the amount delivered at each step will also differ from the programmed amounts, but the total volume of the infusion will be the same as programmed.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/sapphire-multi-therapy-infusio

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.