Product Wiki · Recalls · Health Canada — Medical Devices

Advia and Advia 1200 Chemistry System — recall 50114

Health Canada — Medical Devices · Canada · 2013-10-31 · Type II

As described by Health Canada — Medical Devices: Advia and Advia 1200 Chemistry System - Chemistry Calibrators (2013-09-30)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
50114
Date of alert
2013-10-31
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Quality control (QC) is shifting higher impacting Direct Bilirubin (DBIL_2) with Siemens Chemistry Calibrator lots 090445B and 090445C from previous lots 090445 and 090445A. Observed QC shifts are approx. 7-9% higher with calibrator lots 090445B and 090445C. An urgent field safety notice has been issued by the manufacturer to provide affected customers with reassigned system specific values for the DBIL_2 SSVS for Chemistry Calibrator lot numbers 090445B and 090445C. Manufacturer has determined that the DBIL_2 ssvs currently assigned to these lots may cause an average shift of plus 7-9% in quality control and sample recovery at a level of 3 mg/dL. The urgent field safety notice provides reassigned values for the DBIL_2 ssvs for Chemistry Calibrator lot numbers 090445B and 090445C.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/advia-and-advia-1200-chemistry

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.