Reported risk: Medtronic of Canada Ltd., in consultation with Health Canada, is recalling select lots (see list below) of Paradigm Medtronic MiniMed Insulin Reservoirs, model # MMT-326A and MMT-332A, manufactured by Medtronic MiniMed. These insulin reservoirs are used with Paradigm insulin infusion pumps. Medtronic is recalling these lots due to the potential risk of a leak in the reservoir that may result in under-delivery of insulin. As well, if there is an obstruction in the infusion set, the pump alarm may not be triggered. The under-delivery of insulin may cause hyperglycemia (high blood sugar), and in extreme cases may cause loss of consciousness or death. Medtronic of Canada Ltd is [NJS_FILE:3dd32d88-8ed3-4649-9b5d-64e7ed5b1795:d3a91320-3e1b-43c5-9405-92e1afd6d065] who have purchased products from them of this recall and advising them to check their supplies to see if they have received affected lots. The lot number can be found on the side panel of the reservoir box and on the packaging of each individual reservoir. Patients who have devices from the affected lots or who are concerned about this device should contact Medtronic of Canada Ltd.'s Diabetes Consumer Care line immediately at 1-800-284-4416 or email
[email protected] . Patients who have health concerns related to their use of this device should speak with their health care professional.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.