Reported risk: Siemens Healthcare Diagnostics is informing customers that a rare variant of TSH, identified in a small cluster of patients, is not detected by the siemens assays. As part of an investigation of unexplained discordant results, samples were identified in which the monoclonal antibody used in the reagent failed to detect the TSH molecule. These individuals may have a previously unrecognized, functionally normal TSH variant. The observed rate of occurrence during a 30-month time period was 0.6 x 10-7.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.