Product Wiki · Recalls · Health Canada — Medical Devices

Femur Hook with Support Arm — recall 49561

Health Canada — Medical Devices · Canada · 2013-05-15 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
49561
Date of alert
2013-05-15
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Within the framework of our market surveillance, we have been made aware of one complaint in which a break of a weld of the support arm occurred. Due to the break the complete support arm could fall down. The problem was caused by inaccurate execution of the weld at a supplier. It cannot be excluded that further support arms of the femur hook devices are equipped with a faulty weld which finally could lead to the failure described above.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/femur-hook-support-arm

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.