Immage Immunochemistry Systems Rheumatoid Factor Reagent — recall 49312
Health Canada — Medical Devices · Canada · 2012-12-24 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: During Beckman Coulter Inc (BEC) internal real-time stability testing, some lots of RF reagent showed calibration failure before the currently labeled shelf life. Although some lots have passed real-time stability testing, to be conservative, the expiration dating of all in-date and in-process lots is being shortened.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.