Product Wiki · Recalls · Health Canada — Marketed Health Products

FL14E Rose Bed - Risk of Bed Siderail Collapse - For the Public — recall 49225

Health Canada — Marketed Health Products · Canada · 2013-05-07

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Marketed Health Products
Recall identifier
49225
Date of alert
2013-05-07
Country / jurisdiction
Canada
Risk type
recall
Product category
Medical devices
Product type
Marketed health products
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Stryker Canada, in collaboration with Health Canada, has issued a Notice to Hospitals regarding the Rose Bed, a hospital bed used in healthcare facilities. It was reported that the siderail may appear latched in the upright position when it is not fully locked in place and may suddenly collapse if a pressure is applied on it. This failure has resulted in bruises and/or cuts to patients who are leaning against the siderail when it collapses. Additionally, if this failure occurs, the siderail may not prevent a patient from rolling or falling off the bed.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/fl14e-rose-bed-risk-bed-sidera

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.