Health Canada — Medical Devices · Canada · 2012-12-10 · Type III
As described by Health Canada — Medical Devices: Unicel DXH Slidemaker Stainer Coulter Cellular Analysis system, Software Version 1.5.1.0
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Beckman Coulter Inc. ( BEC ) has confirmed that the system manager on the Unicel DXH Slidemaker Stainer Coulter Cellular Analysis system, software version 1.5.1.0, will not allow the customers to change the time and date backward when following the Instruction for Use ( IFU ). This affects the ability of an operator to make the necessary time change adjustments for the changes from Daylight Savings Time ( DST ) to Standard Time.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.